Поделиться:

Model name | TSTICP7800 Inductively Coupled Plasma Emission Spectrometer |
Voltage | 220V AC±10%\\\\MKJ |
Room temperature | ambient temperature (10-30)℃ |
Relative humidity | (20-80)%RH |
RF frequency | 27.12MHz, coupling efficiency greater than 80% |
RF power stability | ≤ 0.1%;RF frequency stability: ≤ 0.01% |
Light chamber | precise constant temperature 36 ℃ 0.1 ℃, argon driven |
Wavelength range | 165-900nm, full wavelength coverage |
Monochromator | focal length: 400mm, total reflection imaging optical path, Shi Ying prism two-dimensional dispersion system |
echelle grating | 53.6L/mm,63.5 degree shining, the more engraved, the higher the resolution |
Wavelength correction | For each ignition, only Ar spectral line is used to automatically correct the spectral position to ensure the correctness of the analysis wavelength without wavelength correction solution |
stray light | ≤ 2.0mg/L(10000mg/L Ca solution determined at As 188.980nm) |
optical resolution (FWHM) | ≤ 7pm @ 200 nm (resolution and detection limit must be obtained under the same conditions) |
Software performance | It has more than 50000 spectral line libraries, and each spectral line can select at least 30 pixel points for measurement |
With login password protection, multi-level operation permission setting and network security management, and permanent history record saving | |
With visual torch flame observation module | |
Both Chinese and English versions of software | |
computer system | 4GB memory, 500G hard disk, DVD and keyboard and mouse. |
cooling circulating water system | Cooling capacity 1490W, temperature range 5-35 ℃. |
laser printer | Black and white laser printer |
single-phase AC voltage stabilizer | 10KVA, accuracy 1%, input voltage 160V-280V, output voltage 220V. |